The Collimator Data Base File

The collimator data base is stored in a file, called collim.col, that is
located in the special SMC directory. This file, listed below, is a
standard ASCII file in which records of new collimators can be added by
the user. Each records consist of an eight character code that starts
with a '*' character, four real values defining hole diameter, septa
l thickness, hole length and hole shape. The remaining spaces can be used
to add comments. Please be sure to add new collimator data in the same
format as the others are written
COLLIMATOR DATA BASE FOR SIMIND: Note a valid record must start with the character '*'.
All other records are regarded as comments.
If IHAL negative for hex then shape rotated 90 degrees.

------------------------------------------------------------------------------
Code,Diam,Septa,Length,HoleShape,ColType,Fokal,Unused,Unused,Comments
------------------------------------------------------------------------------
;
; GE = GENERAL ELECTRIC
;
*GE-LEGP   0.250   0.030   4.100   3  0  0.000   0.000   0.000    LOW ENERGY GENERAL PURPOSE
*GE-MEGP   0.340   0.140   4.200   3  0  0.000   0.000   0.000    MEDIUM ENERGY GENERAL PURPOSE
*GE-LEHR   0.180   0.030   4.000   3  0  0.000   0.000   0.000    LOW ENERGY HIGH RESOLUTION
*GE-HEGP   0.400   0.320   4.000   2  0  0.000   0.000   0.000    HIGH ENERGY GENERAL PURPOSE
*GE-LEHS   0.340   0.050   3.600   3  0  0.000   0.000   0.000    LOW ENERGY HIGH SENSITIVITY
*GN-LEGP   0.180   0.020   3.150   3  0  0.000   0.000   0.000    NeurCam General Purpose
*GN-LEHR   0.140   0.020   3.150   3  0  0.000   0.000   0.000    NeurCam High Resolution
*GN-LUHR   0.140   0.020   3.850   3  0  0.000   0.000   0.000    NeurCam Ultra-High Resol
*GN-LEHS   0.250   0.020   3.150   3  0  0.000   0.000   0.000    NeurCam High Sensitivity
;
; Millenium VG kameran
;
*GV-LEGP   0.190   0.020   3.500   3  0  0.000   0.000   0.000    LOW ENERGY GENERAL PURPOSE
*GV-LEHR   0.150   0.020   3.500   3  0  0.000   0.000   0.000    LOW ENERGY HIGH RESOLUTION
*GV-EEGP   0.250   0.025   4.500   3  0  0.000   0.000   0.000    EXTENDED LOW ENERGY GENERAL PURPOSE
*GV-MEGP   0.300   0.105   5.800   3  0  0.000   0.000   0.000    MEDIUM ENERGY GENERAL PURPOSE
*GV-HEGP   0.400   0.180   6.600   3  0  0.000   0.000   0.000    HIGH ENERGY GENERAL PURPOSE
*GV-LEUR   0.150   0.020   4.000   3  3  35.00   0.000   0.000    FAN-BEAM LOW-ENERGY ULTRA-HIGH RESOLUTION
*GV-UEUH   0.300   0.190   8.000   3  0  0.000   0.000   0.000    ULTRA-HIGH ENERGY ULTRA-HIGH RESOLUTION
;
;  MPR kamera
;
*GM-LEGP   0.250   0.025   4.300   3  0  0.000   0.000   0.000    LOW ENERGY GENERAL PURPOSE
*GM-LEHR   0.180   0.018   4.100   3  0  0.000   0.000   0.000    LOW ENERGY HIGH RESOLUTION
;
;  GE Infinia
;
*GI-LEHR   0.150   0.020   3.500   3  0  0.000   0.000   0.000    Infinia High Resolution
*GI-LEGP   0.190   0.020   3.500   3  0  0.000   0.000   0.000    Infinia General Purpose
*GI-MEGP   0.300   0.105   5.800   3  0  0.000   0.000   0.000    Infinia MEGP
*GI-HEGP   0.400   0.180   6.600   3  0  0.000   0.000   0.000    Infinia HEGP
*GI-ELEG   0.250   0.040   4.000   3  0  0.000   0.000   0.000    Infinia Extended LEGP
;
; VG = VON GAHLEN COLLIMATORS - HOLLAND
;
*VG-LEUR   0.130   0.020   3.500   3  0  0.000   0.000   0.000    LOW-ENERGY ULTRA HIGH RESOLUTION
*VG-LEHR   0.170   0.020   3.500   3  0  0.000   0.000   0.000    "    "   HIGH RESOLUTION
*VG-LEGP   0.200   0.025   3.500   3  0  0.000   0.000   0.000    "    "   GENERAL PURPOSE
*VG-LEHS   0.240   0.025   3.500   3  0  0.000   0.000   0.000    "    "   HIGH SENSITIVITY
*VG-LEES   0.300   0.030   3.500   3  0  0.000   0.000   0.000    "    "   EXTRA HIGH SENSITIVITY
*VG-LESS   0.300   0.040   2.500   3  0  0.000   0.000   0.000    "    "   SUPER HIGH SENSITIVITY
*VG-LEUS   0.300   0.040   2.000   3  0  0.000   0.000   0.000    "     "  ULTRA HIGH SENSITIVITY
*VG-LETY   0.300   0.020   5.800   3  0  0.000   0.000   0.000    "     "  THYROID
*VG-ME23   0.170   0.020   3.000   3  0  0.000   0.000   0.000    MEDIUM-ENERGY I-123
*VG-ME19   0.240   0.050   3.800   3  0  0.000   0.000   0.000     "     "     190 keV
*VG-ME26   0.450   0.190   3.000   3  0  0.000   0.000   0.000     "     "     260 keV
*VG-ME30   0.340   0.140   4.200   3  0  0.000   0.000   0.000     "     "     300 keV
*VG-ME36   0.450   0.190   6.500   3  0  0.000   0.000   0.000     "     "     360 keV
*VG-METY   0.450   0.190   6.500   3  0  0.000   0.000   0.000     "     "     THYROID
*VG-HE51   0.550   0.250   13.60   3  0  0.000   0.000   0.000               HIGH 511 keV
;
;ENGINEERING DYNAMICS CORPORATION
;
*ED-LEUR   0.140   0.018   2.920   3  0  0.000   0.000   0.000     LOW-ENERGY ULTRA HIGH RESOLUTION
*ED-LEHR   0.140   0.020   2.540   3  0  0.000   0.000   0.000       "    "         HIGH RESOLUTION
*ED-LEGP   0.160   0.020   2.540   3  0  0.000   0.000   0.000       "              GENERAL PURPOSE
*ED-LEMS   0.160   0.023   2.290   3  0  0.000   0.000   0.000       "               MEDIUM SENSITIVITY
*ED-LEHS   0.160   0.023   2.030   3  0  0.000   0.000   0.000       "                 HIGH SENSITIVITY
*ED-LEUS   0.178   0.023   2.030   3  0  0.000   0.000   0.000       "           ULTRA HIGH SENSITIVITY
*ED-MEUS   0.340   0.089   4.950   3  0  0.000   0.000   0.000     MEDIUM-ENERGY ULTRA HIGH SENSITIVITY
*ED-MEHS   0.200   0.102   3.180   4  0  0.000   0.000   0.000       "        "        HIGH SENSITIVITY
*ED-MEHR   0.200   0.102   4.790   4  0  0.000   0.000   0.000       "        "        HIGH RESOLUTION
;
; TO = TOSHIBA MEDICAL SYSTEM
;
*TO-900A   0.236   0.022   4.000   3  0  0.000   0.000   0.000
*TO-901A   0.178   0.017   4.000   3  0  0.000   0.000   0.000
*TO-902A   0.308   0.029   4.000   3  0  0.000   0.000   0.000
*TO-903A   0.136   0.013   4.000   3  0  0.000   0.000   0.000
*TO-921A   0.190   0.025   4.000   3  0  0.000   0.000   0.000
*TO-930A   0.337   0.108   4.000   3  0  0.000   0.000   0.000
*TO-931A   0.260   0.084   4.000   3  0  0.000   0.000   0.000
*TO-908A   1.650   0.370   4.200   2  0  0.000   0.000   0.000
*TO-940A   0.370   0.340   4.000   2  0  0.000   0.000   0.000
;
; NU = NUCLEAR TECHNOLOGIES  HEXARRAY
;
*NU-LESH   0.121   0.018   2.540   3  0  0.000   0.000   0.000     Super High Resolution Low Energy
*NU-LEHR   0.140   0.025   2.540   3  0  0.000   0.000   0.000           High Resolution Low Energy
*NU-EHRC   0.157   0.025   2.540   3  0  0.000   0.000   0.000     Excellent High Resolution Collimator 0.25"
*NU-LEAP   0.178   0.025   2.540   3  0  0.000   0.000   0.000     Optimal Sensitivity vs Resolution All Purpose
*NU-HEGP   0.203   0.025   2.540   3  0  0.000   0.000   0.000     High Efficiency General Purpose Design
*NU-HRDY   0.229   0.025   2.540   3  0  0.000   0.000   0.000     High Resolution "Dynamics"
*NU-HEDY   0.254   0.025   2.540   3  0  0.000   0.000   0.000     High Efficiency "Dynamics"
;
; PR = PRISM 2000/3000 SYSTEMS
;
*PR-LEAP   0.157   0.024   2.540   3  0  0.000   0.000   0.000      Low-Energy All Purpose
*PR-LEHR   0.140   0.018   2.700   3  0  0.000   0.000   0.000      Low-Energy High Resolution
*PR-LEUR   0.140   0.015   3.490   3  0  0.000   0.000   0.000      Low-Energy Ultra-High Res
*PR-ME     0.340   0.086   5.840   3  0  0.000   0.000   0.000      Medium Energy
*PR-HE     0.381   0.173   5.840  -3  0  0.000   0.000   0.000      High Energy
*PR-APFB   0.157   0.024   2.540   3  3  50.00   0.000   0.000      All Purpose Fan-Beam
*PR-HRFB   0.140   0.018   2.700   3  3  50.00   0.000   0.000      High Resolution Fan-Beam
*PR-URFB   0.140   0.015   3.490   3  3  50.00   0.000   0.000      UltraHigh Resol Fan-Beam
;
; SI = SIEMENS MEDICAL SYSTEM
;
*SI-LEAP   0.145   0.020   2.410   3  0  0.000   0.000   0.000     Low-Energy All Purpose
*SI-LEHR   0.111   0.016   2.360   3  0  0.000   0.000   0.000     Low-Energy High Resolution
*SI-ME     0.294   0.114   4.060   3  0  0.000   0.000   0.000     Medium Energy Parallell
*SI-ME+    0.207   0.066   3.284   3  0  0.000   0.000   0.000     Medium Energy Parallell E.CAM+
*SI-HE     0.406   0.200   5.970   3  0  0.000   0.000   0.000       High Energy Parallell
*SI-HE1    0.406   0.200   5.970   4  0  0.000   0.000   0.000       High Energy Parallell
*SI-HE2    0.340   0.200   5.080   3  0  0.000   0.000   0.000       High Energy Parallell
*SI-511    0.340   0.250   7.500   3  0  0.000   0.000   0.000     511 keV Collimator
*SI-512    0.340   0.300   7.500   3  0  0.000   0.000   0.000     511 keV Collimator
*SI-THER   0.250   0.300   5.500   3  0  0.000   0.000   0.000     Therapy Collimator
*SI-DELU   0.100   0.100   1.000   3  0  0.000   0.000   0.000     Test
;
;
; SE = E.CAM SIEMENS MEDICAL SYSTEM
;
*SE-LEHS   0.254   0.036   2.405   3  0  0.000   0.000   0.000     Low-Energy High Sensitivity
*SE-LEAP   0.145   0.020   2.410   3  0  0.000   0.000   0.000     Low-Energy All Purpose
*SE-LEHR   0.111   0.016   2.405   3  0  0.000   0.000   0.000     Low-Energy High Resolution
*SE-LEUR   0.116   0.013   3.580   3  0  0.000   0.000   0.000     Low-Energy Ultra-High Resolution
*SE-LEFB   0.153   0.016   3.500   4  0  0.000   0.000   0.000     Low-Energy Fan-Beam
*SE-ME     0.294   0.114   4.064   3  0  0.000   0.000   0.000     Medium Energy
*SE-HE     0.340   0.200   5.080   3  0  0.000   0.000   0.000     High Energy
*SE-UHE    0.250   0.340   5.050   3  0  0.000   0.000   0.000     Ultra-High Energy
;
;
; SY = SIEMENS MEDICAL SYSTEM SYMBIA COLLIMATORS
;
*SY-LEAP   0.145   0.020   2.405  -3  0  0.000   0.000   0.000     Low-Energy All Purpose
*SY-LEHR   0.111   0.016   2.405  -3  0  0.000   0.000   0.000     Low-Energy High Resolution
*SY-ME     0.294   0.114   4.064  -3  0  0.000   0.000   0.000     Medium Energy Parallell
*SY-HE     0.400   0.200   5.970  -3  0  0.000   0.000   0.000       High Energy Parallell
*SY-HE1    0.400   0.195   5.970  -3  0  0.000   0.000   0.000       High Energy Parallell
*SY-HE2    0.400   0.205   5.970  -3  0  0.000   0.000   0.000       High Energy Parallell
;
;  ADAC Vertex System
;
*AD-LEGP   0.140  0.0180   2.470   3  0  0.000   0.000   0.000     Low-Energy General Purpose
*AD-LEHR   0.140  0.0152   3.280   3  0  0.000   0.000   0.000     Low-Energy High Resolution
*AD-VXGP   0.178  0.0152   4.200   3  0  0.000   0.000   0.000     Low-Energy General Purpose
*AD-VXHR   0.203  0.0152   5.400   3  0  0.000   0.000   0.000     Low-Energy Vertex High Resolution
*AD-UXHR   0.178  0.0152   5.400   3  0  0.000   0.000   0.000     Low-Energy Vertex Ultra-High Resolution
*AD-MEGP   0.295  0.1143   4.800   3  0  0.000   0.000   0.000     Medium Energy General Purpose
*AD-HEGP   0.381  0.1727   6.000   3  0  0.000   0.000   0.000     High Energy General Purpose
*AD-HEHR   0.306  0.1950   6.000   3  0  0.000   0.000   0.000     High-Energy  High Resolution
*AD-511    0.270  0.2300   6.000   3  0  0.000   0.000   0.000     511 keV Collimator
;
;  Trionix System
;
*TR-LEHR   0.140   0.018   2.540   3  0  0.000   0.000   0.000     Low-Energy High Resolution
*TR-LUHR   0.140   0.015   3.492   3  0  0.000   0.000   0.000     Low-Energy Ultra-High Resolution
;
; Picker system
;
*PI-UHE    0.508   0.343   7.700   3  0  0.000   0.000   0.000     Ultra_High-Energy   Collimator
*PI-LEHR   0.122   0.021   2.700   3  0  0.000   0.000   0.000     Low Energy High-Res Collimator
;
------------------------------------------------------------------------------
; Code,Diam,Septa,Length,HoleShape,ColType,Fokal,ClassType,Comments

;===================================================================================================
; Pinhole data!  *** WORK-IN-PROGRESS *** Not publically available yet
;===================================================================================================
; Code,Apertur,Angle,Insert thickness,Holeshape,Coltype,Fokal,Unused,Unused,Comments
;
*PN-TEST FILE:TEST.FIL
*PN-1M40   0.100   80.000   1.000  1  5  29.03   0.000   0.000     Deloar
*PN-2M40   0.200   80.000   1.000  1  5  29.03   0.000   0.000
*PN-3M40   0.300   80.000   1.000  1  5  29.03   0.000   0.000